23
9.2.1 Motion detector activated for presence verification
Symptom
Cause and / or remedy
LED is on during WALK TEST, but no
alarm is triggered
In case of
presence verification
an alarm is only triggered, if
no
motion
has been detected over the programmed period (
time-out
can be selected according to section 5.1.2.1)
WALK TEST has been set, but LED is not
on
Power supply of the motion detector is missing (=> check cable,
connector and power supply)
Power supply not on since two minutes (stabilize time of PIR-motion
detector)
Alarming with announcement „emergency
call activated, sensor 1 activated “
Motion detector has been connected after the unit has been switch on!
Alarming with announcement „emergency
call activated, sensor failure “
Motion detector has been disconnected after the unit has been switch
on!
9.3 Sensor-1-contact
Symptom
Cause and /or remedy
Sensor-1 does not trigger an alarm
by switching on the unit or between two alarms
is not expired
Supervision has been temporarily deactivated (deactivation)
Supervision is deactivated at current position of the
selection switch
(I/II/III) (
section 5.1)
Motion detection is programmed to
verify presence
(
section
5.1.2.1), an alarm is only triggered if
no motion
is detected over a
programmed time.
Test mode => check section 4.2.3
„Alarm due to sensor 1“ is announced, but
the motion detector has been temporarily
deactivated
The power supply of the unit has been cut off for a short period
The
function switch
has been switched to PROG or OFF after
temporarily deactivation.
9.4 Sensor-2-contact
Symptom
Cause and /or remedy
Sensor-2 does not trigger an alarm
by switching on the unit or between two alarms
is not expired
Supervision has been temporarily deactivated (deactivation)
Supervision is deactivated at current position of the
selection switch
(I/II/III) (
section 5.2)
Sensor-2 is activated as an emergency button (
section 5.2.2)
Test mode => check section 4.2.3
„Alarm due to sensor 2“ is announced, but
the motion detector has been temporarily
deactivated
The power supply of the unit has been cut off for a short period
The
function switch
has been switched to PROG or OFF after
temporarily deactivation.
9.5 Sensor-3-contact
Symptom
Cause and /or remedy
Sensor-3 does not trigger an alarm
by switching on the unit or between two alarms
is not expired
Supervision has been temporarily deactivated (deactivation)
Supervision is deactivated at current position of the
selection switch
(I/II/III) (
section 5.3)
Contact is still in alarm position. A new alarm can only be triggered,
after the contact returns to its rest position
Test mode => check section 4.2.3
9.6 Acoustical monitoring
Symptom
Cause and /or remedy
noise is not triggering an alarm
Factory setting supports monitoring of acoustics only on pos. III of
selection switch. (
programming can be altered, see section 5.4)
The unit has been deactivated by DTMF 7 => LED alternatively 4s
on / 4s off
By switching on the unit or after triggering an alarm, the acoustical
monitoring is not active during the
waiting time
of 20 seconds (
section 7.3) => LED is lit constantly during the
waiting time
!
Depending on the position of the selection switch, an alarm is triggered
with a different delay. Every time the noise exceeds the pre set level,
the LED is on. (
section 7.4)