© Leica
TPS-System 1000-2.3.1en
147
CO
IN
ID
PM
FS
SC
OC
IO
AR
RC
EG
CA
CT
BC
DF
SD
TS
IX
The icon
is shown in the lower left area of the
status field whenever the ATR1 mode is active.
MEAS\ MEASURE MODE (GSI)
14:03
ALL
DIST
REC
TARGT Hz0
α
NUM
HELP
LAST
I<>II
PROG
F1
F2
F3
F4
F5
F6
Point no.
:
1
Remark 1
:
-----
Refl.Height :
1.500 m
Hz
:
152°51'36"
V
:
81°24'57"
Horiz.Dist. :
----- m
The error message 565 appears if no reflector is found
after the search routine:
MEAS\ MEAS MODE (GSI)
14:03
ERROR: 565
No prism found or poor
conditions.
Aim accurately at the prism
and retry.
ABORT
CONT
RETRY
F1
F2
F3
F4
F5
F6
Abort the measurement and return to the measure
mode dialog.
Continue with a new distance measurement, without
first re-targeting with the ATR1. This function can be
used if the prism has been located but the required
accuracy has not been attained. This situation could
arise under poor environmental conditions or from a
combination of short measuring distance and unstable
prism.
Repeat the search routine.
The scanning area will be increased by 1/3. At each
repetition the last search area is increased by 1/3.
The centre of the crosshair returns to the initial point
after each search routine if no prism is found.