Safety notes
6
PROVIDO Ref. 10732418 / Version 03
PROVIDO Ref. 10732418 / Version 03 7
CAUTION
Risk of infection.
X
Leave sufficient space around the stand to ensure that
the sterile drape does not come into contact with non-
sterile components.
CAUTION
Risk of damage.
X
Before lifting the microscope make sure that the area
above the stand is clear to avoid collisions with OR
lamps, ceiling etc.
X
Make sure that the movement range is free before you
move the arm with the monitor.
X
Parts of the stand might collide with the ceiling, wall or
other equipment in the environment. Make sure that
the movement range is free before you move the
microscope or stand.
X
Only move the surgical microscope when all brakes are
released.
CAUTION
Risk of overheating of the system.
Covering the air inlet can result in a controlled shutdown of
the system due to overheating.
X
Make sure the air inlet is always kept free.
CAUTION
If the field diameter is greater than the field of view and the
light intensity is too high, uncontrolled tissue heating may
occur outside of the area visible through the microscope.
X
Do not set the light intensity too high.
CAUTION
Destruction of the magnification motor.
X
Only adjust the magnification manually if the
magnification motor is defective.
CAUTION
Destruction of the working distance motor.
X
Only adjust the working distance manually if the
working distance motor is defective.
CAUTION
Damage to the touch panel.
X
Operate the touch panel using your fingers only.
Never use hard, sharp or pointed objects made out of
wood, metal or plastic.
X
Never clean the touch panel using cleaners that contain
abrasive substances. These substances can scratch the
surface and cause it to be become dull.
CAUTION
Danger to the patient due to changes in the user settings.
X
Never change the configuration settings or edit the user
list during an operation.
X
Check fitting and good connection of all parts and
cables during the preparation of the system prior to the
surgery. Not well fitted parts and bad connections can
lead to hazardous situations and system failures.
CAUTION
Danger of skin burns. The lamp insert gets very hot.
X
Check that the cover has cooled before you replace the
lamp.
X
Do not touch the hot lamp insert.
CAUTION
Damage to the PROVIDO surgical microscope during
transportation!
X
Never move the stand in the extended condition.
X
Never roll over cables lying on the floor.
X
Do not transport or store the system in areas with an
elevation angle bigger than 10°.