Specifications
60
PROVIDO Ref. 10732418 / Version 03
PROVIDO Ref. 10732418 / Version 03 61
14.3 PROVIDO floor stand
Type
Floor stand with 6 electromagnetic brakes
Base
700 × 700 mm with four 360° rotating castors with a
diameter of 126 mm each and integrated brakes
Balancing
Manual balancing at microscope carrier and at swing
arm
Floor stand
control unit
The latest electronics control for the continuous
governing of all motor functions and the light
intensity.
Built-in BrightCare Plus safety function for limiting
brightness depending on working distance.
Menu selection based on unique software for user-
specific configuration, with built-in electronic auto-
diagnosis and user support.
Control unit stand Software independent hard key for illumination.
Indicator for Main/backup illumination and
Fluorescence modes.
Open architecture for future software developments.
Light source
Single Xenon arc-lamp as main illumination with LED
lamp as back up.
Dual Xenon arc-lamp illumination system (optional).
Control elements Pistol handle with 10 functions for magnification,
working distance, "All Brakes" button releases
6 brakes, side knob releases selected brake
combinations, motorized side tilt (XY). All buttons
other than "All Brakes" are freely assignable.
Footswitch.
Integrated
documentation
Prepared for integration of video camera system and
digital recording system.
Open architecture
Connectors
Numerous built-in connectors for Video and control
data transfer. Internal power supply 12 VDC, 24 VDC
and AC terminals
Carrier for
monitor
610 mm long and flexible arm with 180° rotation and
inclination to carry optional video monitor
Materials
All solid metal construction
Surface coating
system
Coated with antimicrobial paint
Minimum height In park position: 1975 mm
Range Cantilever Max. 1487 mm
Load
Max. 8.5 kg from microscope dovetail ring interface
Weight
Approx. 350 kg without load
14.4 Ambient conditions
In use
+10 °C to +40 °C
+50 °F to +104 °F
30 % to 95 % rel. humidity
800 mbar to 1060 mbar atmospheric pressure
Storage
–30 °C to +70 °C
–86 °F to +158 °F
10 % to 100 % rel. humidity
500 mbar to 1060 mbar atmospheric pressure
Transport
–30 °C to +70 °C
–86 °F to +158 °F
10 % to 100 % rel. humidity
500 mbar to 1060 mbar atmospheric pressure
14.5 Electromagnetic compatibility
(EMC)
Environment for which the instrument is suitable
Hospitals except for near active HF Surgical Equipment and the RF
shielded room of an ME System for magnetic resonance imaging,
where the intensity of EM Disturbances is high.
Compliance IEC 60601-1-2
Emissions • CISPR 11, Class A, Group 1
• Harmonic Distortion per IEC 61000-3-2 Class A
• Voltage Fluctuation and Flicker per IEC 61000-3-3
Class A, Figures 3-7
Immunity • Electrostatic discharge IEC 61000-4-2:
CD +/– 8 kV, AD +/– 15 kV
• Radiated RF EM Fields IEC 61000-4-3:
80 – 2700 MHz: 10 V/m
• Proximity Wireless fields IEC 61000-4-3:
380 – 5785 MHz: 9 V/m; 28 V/m
• Electrical Fast Transients and
bursts IEC 61000-4-4:
± 2 kV: Power supply lines
• Surges IEC 61000-4-5:
± 1 kV Line-to-line
± 2 kV Line-to-ground
• Conducted disturbances, induced by RF fields
IEC 61000-4-6:
10 V rms
• Rated Power-frequency Magnetic Field IEC 61000-
4-8:
30 A/m
• Voltage dips and interruptions IEC 61000-4-11:
according to IEC 60601-1-2:2014