Functions
14
PROVIDO Ref. 10732418 / Version 03
PROVIDO Ref. 10732418 / Version 03 15
The PROVIDO tilts left:
X
Turn rotary knob C (1) clockwise until the PROVIDO is in balance
with the brakes released.
The PROVIDO tilts right:
X
Turn rotary knob C (1) counter-clockwise until the PROVIDO is in
balance with the brakes released.
X
Press the "AB Free" button.
The button illuminates in green.
Press the ALL-FREE button.
This frees the AB brake.
The PROVIDO tilts backward:
X
Turn rotary knob A (3) counter-clockwise until the PROVIDO is in
balance with the brakes released.
Tilt the PROVIDO by 90°.
The PROVIDO tilts forward:
X
Turn rotary knob B (2) counter-clockwise until the PROVIDO is in
balance with the brakes released.
The PROVIDO tilts backward:
X
Turn rotary knob B (2) clockwise until the PROVIDO is in balance
with the brakes released.
X
Position the PROVIDO horizontally back to the working position.
To avoid sudden tilting when the breaks are released, we
recommend to hold the handle and to slowly let go.
Adjust rotary knob C (1) until the microscope carrier is balanced.
X
Press the "D free" button.
The button illuminates in green.
X
Hold the microscope firmly.