Preparation before surgery
24
PROVIDO Ref. 10732418 / Version 03
PROVIDO Ref. 10732418 / Version 03 25
7.2.2 Releasing the parallelogram
CAUTION
There is a risk of damage to the surgical microscope from
uncontrolled tilting!
X
Firmly hold the handles before triggering the
"All Brakes" function.
X
Grip and turn one handle to release the brakes.
X
At the same time, pull the transport lock (1) and bring it into a
vertical position.
The parallelogram is now released.
If necessary, rebalance the parallelogram.
7.3 Installing optical accessories
WARNING
Risk of injury due to downward movement of the surgical
microscope.
X
Complete all preparations and adjustments to the stand
before the operation.
X
Never change the accessories or attempt to rebalance
the microscope while it is above the field of operation.
X
Before changing accessories, always lock the PROVIDO.
X
Balance the PROVIDO after re-equipping it.
X
Do not release the brakes when the instrument is in an
unbalanced state.
X
Before re-equipping during the operation, first swing
the microscope away from the operating field.
X
Never carry out the adjustment balancing above the
patient.
X
Check fitting and good connection of all parts and
cables during the preparation of the system prior to the
surgery. Not well fitted parts and bad connections can
lead to hazardous situations and system failures.
X
Make sure that the optical accessories are clean and free of dust
and dirt.
7.3.1 Installing a C-mount camera
4
4
2
3
1
3
2
2
2
2
2
X
Remove the cover (1) from the c-mount adapter.
X
Loosen the screw (2).
X
Remove the adapter (3).
X
Set the camera (4) to the adapter (3).
X
Install and adjust camera (4) with adapter (3).
X
Fix the screw (2).
For further information refer to the user manual of the
camera.