Leica LMT260 XY Scanning Stage User Manual
Instrument Setup
30
Operating using the HydraDT control unit
Saving a position of the Leica LMT260 XY scan-
ning stage on one of the push buttons 2-6.
1. Approach the desired position.
2. Press the desired push button for more
than three seconds:
O
The corresponding coordinates (X and Y)
are saved.
O
A successful save is indicated by the corre-
sponding status display flashing three
times.
Saved coordinates are retained until
overwritten by another save operation.
The saved coordinates are the target
coordinates of the last motorized move-
ment. If the stage is unlocked during the save,
what is saved is not the current position, but
instead the last position approached with the
motor!
1 2
3
4
5
6
1
Push button 1
To approach the zero position (center position)
2
Push button 3
To approach or save position 1
3
Push button 4
To approach or save position 2
4
Push button 5
To approach or save position 3
5
Push button 6
To approach or save position 4
6
Push button 2
To approach or save position 5
Five positions can be saved with push
buttons 2 to 6. These positions are not
identical with the saved positions that can be
saved via the LAS or LAS AF software.
Functions of the push buttons
Summary of Contents for LMT260 XY
Page 1: ...Leica LMT260 XY Scanning Stage User Manual ...
Page 3: ...Leica LMT260 XY Scanning Stage User Manual General Notes 3 General Notes ...
Page 10: ...Leica LMT260 XY Scanning Stage User Manual Introduction 10 Introduction ...
Page 16: ...Leica LMT260 XY Scanning Stage User Manual Assembly 16 Assembly ...
Page 22: ...Leica LMT260 XY Scanning Stage User Manual Installation 22 Installation ...
Page 24: ...Leica LMT260 XY Scanning Stage User Manual Instrument Setup 24 Instrument Setup ...
Page 32: ...Leica LMT260 XY Scanning Stage User Manual Service 32 Service ...
Page 36: ...Leica LMT260 XY Scanning Stage User Manual Specifications 36 Specifications ...