Remove the transport anchor
1
. Remove the transport anchor screws.
Secure the axes with your fingers for
the following steps, so as not to jam the
axes by the movements of the Leica LMT260 XY
scanning stage.
Keep the safety screws; they are needed
for retransporting the stage. Do not
transport the stage without first installing the
safety screws.
Mount the Leica LMT260 XY scanning stage
1. Place the Leica LMT260 XY scanning stage
to fit precisely on the adapter.
2. Gently tighten both front M4x12 Allen
screws with a 3 mm Allen key.
3. Gently tighten both rear M4x12 Allen
screws with a 3 mm Allen key.
4. To guarantee a high level of precision for
the stage, tighten the four screws crosswise
exactly 1.6 Nm using a torque wrench.
Mount the glass stage plate (optional)
1. Mount the glass stage plate (optional)
or the Leica Live on Stage sample holder
(optional) as shown.
Leica LMT260 XY Scanning Stage
1
2
2
3
3
1.6 Nm
SW 3
1
1
1
Summary of Contents for LMT260 XY
Page 1: ...Leica LMT260 XY Scanning Stage User Manual ...
Page 3: ...Leica LMT260 XY Scanning Stage User Manual General Notes 3 General Notes ...
Page 10: ...Leica LMT260 XY Scanning Stage User Manual Introduction 10 Introduction ...
Page 16: ...Leica LMT260 XY Scanning Stage User Manual Assembly 16 Assembly ...
Page 22: ...Leica LMT260 XY Scanning Stage User Manual Installation 22 Installation ...
Page 24: ...Leica LMT260 XY Scanning Stage User Manual Instrument Setup 24 Instrument Setup ...
Page 32: ...Leica LMT260 XY Scanning Stage User Manual Service 32 Service ...
Page 36: ...Leica LMT260 XY Scanning Stage User Manual Specifications 36 Specifications ...