background image

4

4.4 Examining Specimen at High Power

When examining a specimen at high magnification, position the slide

so the feature is centered in the field of view. If the feature you want

to observe appears to the left of the field of view, move the slide left

to center it. If the feature you want to observe appears to the right of

the field of view, move the slide to the right.

By rotating the nosepiece, bring the next higher power objective into

viewing position. The objective is properly aligned when the nosepie-

ce clicks into place.

4.5 Oil Immersion Technique

To utilize the full numerical aperture of an immersion objective (with

inscription “oil”), oil is added to both the objective and specimen using

the following procedure:

1. Focus on the specimen with a lower power objective (40x).

2. Put a drop of immersion oil on the specimen slide.

3. Turn the nosepiece to bring the immersion objective into the

light path, and focus with the fine adjustment knobs.

Care should be taken to prevent bubbles from forming in the oil, sin-

ce they will deteriorate the lens performance. Simply reapply the

immersion oil if bubbles form.

IMPORTANT: 

After each use of the 100x objective, use a lens tissue

or soft cloth to wipe off all traces of oil from the objective and the sli-

de.

5.0 Care of the 

Microscope

5.1 General

It is important to keep all optical components clean to obtain the best

optical performance. Keep a dust cover on the microscope when it is

not in use.

The objectives are easily subjected to dirt, dust and oil. Whenever

lack of contrast, cloudiness or poor definition is encountered, careful-

ly check the condition of the front objective lens with a magnifier. To

assure all objective lenses are clean, periodically examine them with

a magnifying glass.

If an optical surface becomes badly coated with dust or dirt, use an

air-filled syringe or soft brush to remove it before wiping the surface

clean.

IMPORTANT: 

Clean optical surfaces with a lint-free cloth, lens tissue

or a cotton swab moistened with methanol or commercially availa-

ble glass cleaner. Avoid excessive use of solvents. Interior surfaces

can be loosened or badly damaged if solvents seep in around the len-

ses.

The higher magnification objectives use a small front lens. The surfa-

ce of this front lens can be cleaned with a small cotton swab. Dampen

the cotton tip with methanol and squeeze it dry. Wipe the front lens

lightly. Make sure that the cotton tip contacts the lens surface. Check

the objective with a magnifier after the cleaning.

IMPORTANT: 

If it is necessary to remove the viewing body,  DO NOT

TOUCH the outer surface on the underside of the viewing body. Finger

prints on this surface will reduce the image clarity. This lens can be

cleaned in the same manner as objectives and eyepieces.

5.2 Mechanical Maintenance

All non-optical components of the Leica BM E Compound Microscope

can be cleaned with soap and water. Avoid using all other harsh sol-

vents or acetone-based products when cleaning.

The Leica BM E Compound Microscope requires only routine main-

tenance. Neither the focusing mechanism nor the mechanical stage

require lubrication. Contact your Leica customer service representa-

tive for more information.

English

Summary of Contents for BM E

Page 1: ...Leica BM E Instruction Manual Gebrauchsanleitung Mode d emploi Manual de instrucciones Manuale d istruzione M I C R O S Y S T E M S...

Page 2: ......

Page 3: ...and illumination system mirror or in base illuminator 2 Viewing Body monocular binocular or monocular teaching as ordered 3 Objectives as ordered 4 Eyepiece s as ordered 5 Immersion Oil as ordered 6 D...

Page 4: ...tial focusing and viewing The lower the objective power the greater area of specimen surface will be seen in the field of view Before placing a specimen on the stage rotate the nosepiece until the low...

Page 5: ...y subjected to dirt dust and oil Whenever lackofcontrast cloudinessorpoordefinitionisencountered careful ly check the condition of the front objective lens with a magnifier To assure all objective len...

Page 6: ...t power 20VA 220 240VAC 10 50 60Hz input power 45VA Range of Environmental Conditions Indoor use only Temperatures 5 C to 40 C 41 F to 104 F Maximum Relative Humidity 80 for temperatures up to 31 C 87...

Page 7: ...icaBMEistalsbereitskonfiguriertesModellerh lt lich berpr fen Sie das Vorhandensein der Teile anhand der nach folgenden standardm igen Ausr stungsliste Diestandardm igenKomponentendesMikroskopesLeicaBM...

Page 8: ...sundgleichm igausgeleuchtetesGesichtsfeldzuerhal ten 4 2 Vorbereitung des Objekttisches Objektive mit einem niedrigeren Vergr erungsfaktor haben eine gr ere Tiefensch rfe und werden in der Regel f r a...

Page 9: ...her Vergr erung alle lspuren vom Objektiv und dem Objekttr ger mit einem Linsentuch bzw einem weichen Tuch ab 5 0 Pflege des Mikroskops 5 1 Allgemeine Hinweise Es ist wichtig alle Optikkomponenten f r...

Page 10: ...40 C Max relative Luftfeuchtigkeit 80 bei Temperaturen bis zu 31 C mit linearem Abfall auf 50 relative Luftfeuchtigkeit bei 40 C Installationskategorie II berspannungskategorie Verschmutzungsgrad 2 H...

Page 11: ...le m canisme de mise au point la platine le condenseur ou le diaphragme disque le revol ver d objectifs et le syst me d clairage miroir ou illuminateur dans la base 1 Modules d observation monoculair...

Page 12: ...ent utilis s pour la mise au point et l observation initiales Plus faible est la puissance de l objectif plus grande est la surface de l chantillon apparaissant dans le champ de vision Avant de placer...

Page 13: ...s d huile de l ob jectif et de la lamelle 5 0 Entretien du microscope 5 1 G n ralit s Il est important que les composants optiques soient toujours propres pour maintenir une bonne performance optique...

Page 14: ...a 10 50 60 Hz alimentation 45 VA Plage et conditions environnementales Pour utilisation l int rieur uniquement Temp rature 5 C 40 C 41 F 104 F Humidit relative maximum 80 pour les temp ratures jusqu...

Page 15: ...del Microscopio Compuesto BM E Leica son 1 Pedestal incluye brazo de soporte mecanismo de enfoque platina condensador o diafragma de disco portaobjetivo y sistema de iluminaci n espejo o iluminador en...

Page 16: ...i n de la platina Los objetivos de menor potencia tienen mayor profundidad de cam po y se usan generalmente para el enfoque y visualizaci n iniciales Cuanto menor es la potencia del objetivo mayor es...

Page 17: ...pa o suave para limpiar todas las trazas de aceite del objetivo y del portaobjetos 5 0 Cuidado del microscopio 5 1 Generalidades Es importante mantener todos los componentes pticos limpios para obten...

Page 18: ...45 VA Rango de condiciones medioambientales S lo para uso en interiores Temperaturas 5 C a 40 C 41 F a 104 F Humedad relativa m xima 80 para temperaturas hasta 31 C 87 8 F disminuyendo linealmente a...

Page 19: ...roscopio Leica BM E sono 1 Stativo include braccio di sostegno meccanismo di messa a fuoco tavolino portaoggetti condensatore o diaframma a disco revolver e sistema d illuminazione specchietto o illum...

Page 20: ...o e pi uniformemente illuminato possibile 4 2 Preparazione del tavolino portaoggetti Gli obiettivi a minor ingrandimento posseggono una maggiore pro fondit di campo e generalmente sono usati per la me...

Page 21: ...per immersione IMPORTANTE Dopo ciascun uso dell obiettivo 100x pulire ogni trac cia d olio dall obiettivo e dal vetrino usando una velina per lenti o un panno morbido 5 0 Manutenzione del microscopio...

Page 22: ...5VA Gamma di Condizioni ambientali Da usarsi solo in ambienti interni Temperatura da 5 Celsio a 40 Celsio da 41 F a 104 F Massima Umidit Relativa 80 per temperature fino a 31 Celsio 87 8 F diminuendo...

Page 23: ...22...

Page 24: ...olique de d veloppement continu nous nous r servons le droit de modifier les sp cifications sans pr avis En vista de nuestra pol tica de desarrollo continuo nos reservamos el derecho de modificar las...

Reviews: