Instructions for Use V 2.3 – 10/2012
14
Instructions on how to install the instrument are provided in this chapter. A
diagram and description of components is also given. Finally, the procedure
for replacement of the fume filter is outlined.
4.1
Site requirements
The AutoStainer XL requires a solid bench of dimensions 1.090 mm long and
670 mm deep. The instrument must be located within 3 meters of a tap and
drain.
The power requirements for the AutoStainer XL are:
8 amps: at 110 volts
4 amps: at 240 volts
The voltage selector and other internal components are set by the manu-
facturer to suit the country of sale.
The voltage selector setting must not be altered by the user.
The AutoStainer XL requires connection to a laboratory water tap with mains
pressure fitting.
4.2
Connection
4.2.1 Power
•
Connect the power cord to the mains outlet (
(28)
page 11).
•
Connect the power jumper cable to the power supply outlet (
(30)
page
11) and to the power supply inlet (
(32)
page 11).
Refer to Figure 10 on page 11.
4.
Installation