Leica AutoStainer XL – Automated slide stainer
5.8
Rack movement times
You can set the rack agitation, withdrawal and entry times to suit your run
time and agitation requirements. Refer to Appendix 1 for the allowable ranges.
To view or change the settings:
1. Press
[F3]
SetUp
from the
Main Menu.
2. Press
[F3]
Mov
e.
The current settings are displayed shown as the seconds taken for each
movement, i.e. agitation cycle time (Dip), rack withdrawal time (Up) and
rack entry time (Down).
3. To change any of the values, press
[F1]
Dip
,
[F2]
Up
or
[F3]
Down
to
position the cursor beneath the appropriate value.
4. Use the
and
keys to alter the setting.
5. Repeat steps 3 and 4 as required.
6. Press
[F4]
Return
to return to the
SetUp
screen.
7. Press
[F4]
Return
to return to the
Main Menu
.
5.9
Staining
This section provides a guide to staining slides.
The AutoStainer XL can accept slide racks whenever the load station is empty
and stain them according to the program selected for each rack. Different
programs may be used simultaneously provided they are compatible. To
check whether programs are compatible, refer page 26.