11.2 Cooling capacity
Tab. 14
Device
Cooling
capacity (20°C)
Cooling
capacity
(10 °C)
Cooling
capacity (0 °C)
Cooling
capacity
(-10 °C)
Cooling
capacity
(-20 °C)
kW
kW
kW
kW
kW
VC 1200 (W)
1.20
1.00
0.70
0.40
0.14
VC 2000 (W)
2.00
1.50
1.06
0.68
0.38
VC 3000 (W)
3.00
2.40
1.68
0.95
0.45
VC 5000 (W)
5.00
3.90
2.75
1.70
0.90
VC 7000 (W)
7.00
5.30
3.70
2.40
1.30
VC 10000 (W)
10.00
7.60
5.30
3.50
2.00
The cooling capacity is measured for a specified tem-
perature of the heat transfer liquid. These temperature
values are shown in brackets. The ambient temperature
for the measurement is 20 °C; ethanol was used as heat
transfer liquid. The cooling water temperature is 15 °C
and the cooling water differential pressure is 3 bar for
the measurement of water-cooled devices.
All water-cooled Variocool circulation chillers have the following
cooling water connection:
n
¾
" external connection thread
11.3 Refrigerant and filling quantity
The device contains fluorinated greenhouse gases.
Tab. 15
Unit
VC 1200
VC 2000
VC 1200 W
VC 2000 W
Refrigerant
---
R-449A
R-449A
R-449A
R-449A
Maximum filling weight
kg
0.50
0.58
0.50
0.58
Cooling water connection
Technical Data
Variocool
95