Code
English output
Description
024
SW Pump 2 old
Software version of pump 2 too old
025
SW Pump 3 old
Software version of pump 3 too old
026
SW HTC old
Software version of high temperature cooler too old
027
SW Ext. Pt100 old
Software version of external Pt100 too old
028
SW Ethernet old
Software version of Ethernet too old
029
SW EtherCAT old
Software version of EtherCAT too old
033
Clock wrong time
Internal clock defective; battery power supply was/is interrupted
(insert new battery)
034
Tset: Prog. is running
Setpoint was changed while the programmer is running.
041
Wrong mains voltage
Incorrect mains voltage setting
042
No VC type
Device type not configured
043
No VC voltage
Mains voltage not configured
050
Level very low
Level too low, top-up heat transfer liquid
051
Level high
Level too high (fill level of the heat transfer liquid too high, risk of
bath overflow)
055
CAN buff. overflow
Buffer overflow for CAN reception
8.6 Warnings – safety system
All warnings of the safety system start with the prefix 1.
The prefix is followed by two other digits. These digits
are listed in the following table.
Code
English output
Description
101
CAN receive overflow
Overflow during CAN reception
102
Watchdog Reset
Watchdog reset
103
Heating not correct
Heaters have different outputs
104
Heat 1 failed
Heater 1 defective
105
Heat 2 failed
Heater 2 defective
106
Heat 3 failed
Heater 3 defective
107
Invalid Parameter
Inadmissible parameter in memory
108
CAN system
Problem during internal data exchange
109
Unknown Modul
Unknown module connected
110
SW Control too old
Software version of control system too old
Faults
Variocool
87
Summary of Contents for VC 1200
Page 47: ...Fig 13 Menu structure part 1 Operation Variocool 47...
Page 48: ...Fig 14 Menu structure part 2 Menu structure for Graph Clock and Standby Operation Variocool 48...
Page 103: ...General Variocool 103...
Page 110: ......
Page 111: ......