11.6 Equipment, voltage-independent
Devices
Sound insulation
Outdoor installation
Insulation of the cooling
water hydraulic system
VC 5000
X
X
---
VC 7000
X
X
---
VC 10000
X
X
---
VC 1200 W
---
---
X
VC 2000 W
---
---
X
VC 3000 W
---
---
X
VC 5000 W
X
---
X
VC 7000 W
X
---
X
VC 10000 W
X
---
X
11.7 Line fuse
Alternating current
VC 1200 (W)
VC 2000 (W)
VC 3000 (W)
230 V; 50 Hz
T10 A
T16 A
T16 A
with heater
T16 A
T16 A
T16 A
200 V; 50/60 Hz
T16 A
T16 A
T16 A
with heater
T16 A
T16 A
T16 A
208–220 V; 60 Hz
T16 A
T16 A
T16 A
with heater
T16 A
T16 A
T16 A
Three-phase current
VC 5000 (W)
VC 7000 (W)
VC 10000 (W)
400 V; 3/N/PE~50 Hz
T16 A
T16 A
T16 A
with heater
T16 A
T16 A
T16 A
208-220 V; 3/PE~60 Hz
T16 A
T20 A
T25 A
with heater
T16 A
T20 A
T25 A
200 V; 3/PE~50/60 Hz
T16 A
T20 A
T25 A
with heater
T16 A
T20 A
T25 A