104 / 137
Proline Kryomats
8.3.6
Read commands (Data requested from the thermostat)
Command
Explanation
IN_PV_00
Read bath temperature (outflow temperature).
IN_PV_01
Indication of the controlled temperature (internal / external Pt / external Analogue / external
serial).
IN_PV_03
Read external temperature TE (Pt100).
IN_PV_04
Read external temperature TE (Analogue input).
IN_PV_05
Read bath level.
IN_PV_10
Read bath temperature (outflow temperature) in 0.001 °C.
IN_PV_13
Read external temperature TE (Pt100) in 0.001 °C.
IN_SP_00
Read temperature setpoint.
IN_SP_01
Read current pump power level.
IN_SP_02
Read cooling operation mode (0 = OFF / 1 = ON / 2 = AUTOMATIC).
IN_SP_03
Read current overtemperature switch-off point.
IN_SP_04
Read current outflow temperature limit TiH.
IN_SP_05
Read current outflow temperature limit TiL.
IN_PAR_00
Read control parameter Xp.
IN_PAR_01
Read control parameter Tn (181 = OFF).
IN_PAR_02
Read control paramete Tv.
IN_PAR_03
Read control parameter Td.
IN_PAR_04
Read control parameter KpE.
IN_PAR_05
Read control parameter TnE (980 = OFF).
IN_PAR_06
Read control parameter TvE (0 = OFF).
IN_PAR_07
Read control parameter TdE.
IN_PAR_09
Read value of correction limitation
IN_PAR_10
Read control parameter XpF.
IN_PAR_14
Interrogation of the setpoint offset.
IN_PAR_15
Read control parameter PropE.
IN_DI_01
State of contact input 1: 0 = open/ 1 = closed.
IN_DI_02
State of contact input 2: 0 = open/ 1 = closed.
IN_DI_03
State of contact input 3: 0 = open/ 1 = closed.
IN_DO_01
State of Contact output 1:
0 = make-contact open/ 1 = make-contact closed.
IN_DO_02
State of Contact output 2:
0 = make-contact open/ 1 = make-contact closed.
IN_DO_03
State of Contact output 3:
0 = make-contact open/ 1 = make-contact closed.
IN_MODE_00
Keyboard Master: 0 = free / 1 = locked. .
IN_MODE_01
Control: 0 = internal / 1 = external Pt100 / 2 = external Analogue / 3 = external Serial / 5 =
external Ethernet / 6 = external EtherCAT..
IN_MODE_02
Standby operation: 0 = Device ON / 1 = Device OFF.
IN_MODE_03
Keyboard Command remote control: 0 = free / 1 = locked.
IN_MODE_04
Setpoint offset source: 0=normal / 1=external Pt / 2=external analogue / 3=external serial.
TYPE
Read equipment type (response e.g. “RP 3090”).
VERSION_R
Read software type of control system.
VERSION_S
Read software type of protection system.