If the field in the column "Tolerance" has a
black background, the entry mode for the
"Temperature tolerance" is obtained by
pressing the
key. It defines how accurately
the end of segment temperature is to be
obtained before the next segment is processed.
A tolerance which is selected too small can
stop the next segment from being started
according to plan.
Set the temperature tolerance and confirm
with
.
Continue with
to the entry field "Pump".
If the field in the column "Pump" has a black
background, the entry mode for the
Pump
level
is obtained by pressing the key
.
With
or
select Pump Level 5 – 8 or
“-------” and confirm with
.
“------” stands for "no change to previous
segment", i.e. when “------” is present in all
fields, the pump level always retains the start
setting or the setting before the program start.
Continue with
to the field "Out 1", "Out 2"
or "Out 3".
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Temp. tolerance (0=off):
10.
00
Min: 0.00°C Max:450.00°C
Pump Menu
End
T
set
T
fix
Level 8
Level 7
Level 6
Level 5
Level 4
Level 3
Level 2
Level 1
-------
Pump level