Proline Kryomats
123 / 137
Table 2
RP 3090 C
RP 3090 CW
RP 4090 C
RP 4090 CW
Operating temperature - ACC range
°C
-90 – 200
Ambient temperature range
°C
5 – 40
Relative humidity
maximum relative humidity 80 % for temperatures up to 31 °C, decreasing
linearly to 50 % relative humidity at 40 °C
Device distance to the surroundings
cm
50
20
50
20
Temperature range for storage
°C
-20 – 44
the condenser must be completely emptied by a water-cooled device (
Setting resolution
°C
0.1 / 0.01 (Master); 0.01 (Command)
Display resolution
°C
Master: 0.01 Command: 0,1 / 0,01 / 0,001
Display accuracy
±0.2 °C can be calibrated additively (
Section 1.3 last Point)
Temperature stability
@ -10 °C with ethanol
K
±0.05
Safety equipment
Class
III, FL suitable for flammable and non-flammable liquids
Cooling
Air
Water
Air
Water
Cooling water consumption:
temperature 15 °C, pressure 3 bar
L/h
700
700
Water-cooling connections
(DIN EN 10226-1)
inch
G ¾”
G ¾”
Heater power 400 V
kW
maximum 3.5
Heater power 208 V
kW
maximum 3.0
Heater power 200 V
kW
maximum 2.8
C
ool
in
g p
ow
er
at
2
0
°C
t
am
b
@ b
at
h
tem
p.
(P
um
p L
ev
el
6)
with heat
transfer oil
200 °C
kW
3.0
4.0
3.0
4.0
with ethanol
20 °C
kW
3.0
4.0
3.0
4.0
0 °C
kW
2.9
3.7
2.9
3.7
-20 °C
kW
2.5
3.1
2.5
3.1
-30 °C
kW
2.3
2.7
2.3
2.7
-40 °C
kW
2.0
2.0
2.0
2.0
-50 °C
kW
1.6
1.6
1.6
1.6
-60 °C
kW
1.3
1.3
1.3
1.3
-70 °C
kW
0.8
0.8
0.8
0.8
-80 °C
kW
0.5
0.5
0.5
0.5
-90 °C
kW
0.15
0.15
0.15
0.15