Proline Kryomats
105 / 137
Command
Explanation
VERSION_B
Read software type of Command.
VERSION_T
Read software type of cooling system.
VERSION_A
Read software type of analogue module.
VERSION_V
Read software type of RS 232/485 module.
VERSION_D
Read software type of digital module.
VERSION_M_0
Read software type of solenoid valve (cooling water).
VERSION_M_1
Read software type of solenoid valve (automatic refill).
VERSION_M_3
Read software type of solenoid valve (shut-off valve 1)
VERSION_M_4
Read software type of solenoid valve (shut off valve 2).
VERSION_M_5
Read software type of high temperature cooler.
STATUS
Read equipment status 0 = OK, -1 = error.
STAT
Read error diagnosis response:
XXXXXXX
X = 0 no error, X = 1 error.
1. char = error.
2. char = Alarm.
3. char = Warning.
4. char = over temperature.
5. char = low-level error.
6. char = high-level error (at adjustment alarm).
7. char = no external control variable.
RMP_IN_00_XXX
Read a program segment XXX
(response: e.g. 030.00_010.00_005.00_001.00
set point temperature 30.00 °C,
time = 10 min, tolerance = 5.00 °C, pump level = 1).
RMP_IN_01
Read the current segment number.
RMP_IN_02
Read the set number of program runs.
RMP_IN_03
Read the current program run.
RMP_IN_04
Read the program to which further instructions apply.
RMP_IN_05
Read which program is currently running (0 = none).
LOG_IN_00_XXXX
Query a measuring point XXXX from data logger
(Reply: e.g. 020.00_021.23_030.50 => set point temperature = 20.00 °C, bath
temperature = 21.23 °C, external temperature = 30.5 °C).
LOG_IN_01
Read all measuring points from data logger
As a difference to the command “LOG_IN_00”, a tabulator is used here as separator instead
of ,_’ . The measuring points are separated by CR and LF. The end is marked by CR LF CR
LF.
LOG_IN_02
Read the start time from data logger
(Reply: e.g. 20_14_12_20
day 20, time 14:12:20).
LOG_IN_03
Read the acquisition interval from the data logger (Reply in seconds).