Press
Q
to select
CONTINUE
and go to the
HI
STD
. Scroll to and select
SCAN AGAIN
if some
mistake was made in scanning.
51. H2O
DONE LO STD
*CONTINUE
SCAN AGAIN
Use the same procedure to enter concentration
of
HI STD
as was used for entering the
LO STD
.
Once the number is complete, select =
I
to end
number and verify number in next menu.
51. H2O
*HI STD
=
1
2
3
:
•
9
Press the
Q
button to select
KEEP
and accept the
displayed concentration of the
HI STD
.
This menu allows the user to confirm that the
correct number has been entered. If
CHANGE
is
selected a number must be entered.
51. H2O
HI STD=10.00
*KEEP
CHANGE
Place reacted sample of high concentration
standard (
HI STD
) in the colorimeter chamber.
Press the
Q
button to scan sample. The
colorimeter scans the sample and the next display
appears.
51. H2O
*SCAN HI STD
SMART COLORIMETER
43