TESTING WITH THE %T/ABS TESTS
Scroll to %T/ABS with the
t
button in the
TESTING MENU
.
Push
Q
button to select %T/ABS.
TESTING MENU
*%T / ABS
ALL TESTS
Use
t
to scroll to desired wavelength.
%T / ABS
*43. %T/ABS 420
44. %T/ABS 460
45. %T/ABS 510
46. %T/ABS 530
47. %T/ABS 570
48. %T/ABS 605
74. END OF LIST
Push
Q
button to select wavelength.
(Here 530nm is selected as an example.)
The SMART Colorimeter is ready to scan.
%T / ABS
*46. %T/ABS 530
47. %T/ABS 570
48. %T/ABS 605
74. END OF LIST
Insert blank into colorimeter chamber and push
Q
button to scan blank. The filter will be
positioned and the blank scanned.
43. %T/ABS 530
*SCAN BLANK
SCAN SAMPLE
NEXT TEST
PREVIOUS TEST
NOTE: For the %T/abs tests, a clear colorless water blank should be
used.