To access
EDIT USER TESTS
menu, the
colorimeter must be
OFF!
Press and hold down
the
Q
button (with colorimeter off), then press
the
USE
button (to turn the colorimeter on).
This menu will be displayed:
Scroll to and select the USER TEST to be
changed. (In this example
51. USER TEST 3
).
EDIT USER TESTS
*49. USER TEST 1
50. USER TEST 2
51. USER TEST 3
52. USER TEST 4
:
74. END OF LIST
NOTE:
This menu allows a test to be renamed,
a wavelength to be selected, two reacted
standards to be scanned and the numerical
format of the displayed test result to be selected.
After editing any one of these menu choices the
display will return to this menu. Any menu
choice can be edited at any time by selecting it.
The normal procedure would be to start with
NAME THE TEST
, then
SELECT FILTER
, then
NEW CALIBRATION
, and then
FORMAT
RESULT
.
EDIT USER TEST 3
*NAME THE TEST
SELECT FILTER
NEW CALIBRATION
FORMAT RESULT
NAMING THE TEST
A NAME can be up to 10 characters long. The menu choices for each
character are 26 letters A to Z, ten numerals 0 to 9, a space, a dash, a
decimal point, and a =
I
as a terminator. Selecting the terminator indicates
the end of the name and stores the name. The terminator, =
I
, is the first
menu choice since it is the one character that will always have to be
selected. It is before the letter “A”. DO NOT CHOOSE THE =
I
UNTIL
THE NAMING OF THE TEST HAS BEEN COMPLETED. The
terminator should be selected following the end of the name. It must be
the last character selected.
SMART COLORIMETER
35