EDIT SEQUENCES
The EDIT SEQUENCES mode allows any of the three alterable test
sequences (
SEQUENCE 1
,
SEQUENCE 2
, and
SEQUENCE 3
) in the
TESTING MENU
to be edited. This feature allows a sequence to be set up
which contains the tests which are used frequently. The order of the tests
in a sequence can be arranged to suit the needs of the user. Any
combination, and any order of tests from the
TEST LIBRARY
, may be
placed into these sequences. The only restriction is that the total number
of tests in a sequence does not exceed the total number of tests (74) in
the
TEST LIBRARY
.
To access the
EDIT SEQUENCES
the colorimeter
must be turned off!!
Press and hold down the
EDIT
button (with the colorimeter off) then
press the
USE
button (to turn the colorimeter
on). This menu will appear:
EDIT SEQUENCES
*EDIT SEQUENCE 1
EDIT SEQUENCE 2
EDIT SEQUENCE 3
TEST LIBRARY
The first three menu choices in
EDIT SEQUENCES
allow the three
alterable sequences to be edited.
TEST LIBRARY
will allow scrolling
through a list of all the available tests.
ADDING OR DELETING TESTS
There are two ways to alter a sequence -
INSERT BEFORE
and
DELETE
.
INSERT BEFORE
is used to add a new test to a sequence and to place the
new test before an existing test in a sequence.
DELETE
is used to remove an existing test from a sequence.
Below is a step by step example of how to
EDIT SEQUENCE 3
starting
from the
EDIT SEQUENCES
menu. The
29. PH CPR
test will be removed
and
1. ALUMINUM
will be added to the sequence in the position
previously occupied by the pH test.