background image

No.:GC-JS-21                                Version: 01/00                                       

Effective date

2020-07-22

 

 

30 

E62 

220

heating jacket 

temperature> 190 

 

1. The connector of the 
temperature sensor of the 
heating ring is not in good 
contact 
2. The circuit board is damaged 

Re-made the connector of the 
temperature sensor of the heating 
jacket 
Replace the circuit board 

E63 

230

heating jacket 

temperature> 160 

 

1. The connector of the 
temperature sensor of the 
heating jacket is not in good 
contact, 
2. The circuit board is damaged 
3. The pump is damaged 

Re-made the connector of the 
temperature sensor of the heating 
jacket 
Replace the circuit board 
Replace the water pump 

10 

E2 

Overpressure in 
sterilization chamber: 
134 

  pressure exceeds 

235kpa, 121 

  program 

pressure exceeds 135kpa 

No  vacuum  can  be  drawn,  and 
more cold air remains 

Test vacuum pump, vacuum 
selection valve 

11 

E41 

After 8 minutes in the 
preheating phase, the 
temperature of the 
heating jacket <100 

 

heating jacket failed 

Check the thermostat of heating 
jacket and the related thermostat 

12 

E42 

The steam generator 
temperature <110 

 

after 8 minutes in the 
preheating stage 

Heating bar in steam generator 
failed 

Check the thermostat of heating 
bar and the related thermostat 

13 

E5 

After 10 minutes of 
sterilization phase, the 
exhaust failed to reach 
20kpa

 

Drain piping blocked 

Clean and sterilize indoor filters 
Check the drain solenoid valve, 
vacuum selection valve 

14 

E6 

The door switch is 
disconnected when the 
program is running 

The  door  switch  sensor  is 
displaced,  the  micro  switch 
cannot be closed 

Reassemble the door switch 

15 

E7 

Local atmospheric 
pressure is less than 
<70KPa; 

1.  The  stored  pressure  value  is 
wrong, 
2.  The  ambient  air  pressure  is 
too low 

Corrected atmospheric pressure 
Check if ground elevation is 
detected above 2500 meters 

16 

E8 

During the rising process, 
the pressure is added less 
than 3kpa in 1 minute 

1. unable to inject water; 
2. The heating bar of the steam 
generator is damaged 
3. The door seal is damaged 

Observe whether the water 
storage tank has enough water and 
check the water level sensor; 
Check the pump and inlet 
solenoid valve; 
Check the door seal 

17 

E9 

Low pressure during 
sterilization: 134 

 

pressure is lower than 
100kpa, 121 

  program 

pressure is lower than 
200kpa 

1. too much loading loaded; 
2. unable to enter the water 

1. Reduce the number of loads; 
2. Observe whether the water 
storage tank has enough water and 
check the level switch; 
3. Check the water pump and inlet 
solenoid valve 

18 

E10 

Door can not be locked at 
the beginning of cycle 

Door lock failure 

Check electromagnet and circuit 
board 

19 

E11 

The door lock cannot be 
opened at the end of the 
program 
Door lock is open during 
operation 

1.

 

The electronic lock is stuck 
and damaged 

2.    The circuit board is 
damaged 

Check the electronic lock and 
circuit board 

Summary of Contents for Autoclave LFSS03AA

Page 1: ...0 Manufacturer Ningbo Haishu Life Medical Technology Co Ltd No 1 Jinghui Road Hengjie Haishu Ningbo 315181 China P R C EC Representative Caretechion GmbH Niederrheinstr 71 40474 Duesseldorf Germany Pl...

Page 2: ...No GC JS 21 Version 01 00 Effective date 2020 07 22 2 Revision History Revised chapter Revised content Revised date Revisor Revised number All New document 2020 7 22 Kejie Zheng V1 0...

Page 3: ...stalltion 12 4 Adjusting 16 5 Introduction of preset programs 17 6 User interface description 19 Instructions of operation 25 8 Errors list 29 9 Maintenance 32 10 Quality assurance 37 11 Accessories 3...

Page 4: ...manual must be transferred or handed over as part of the whole equipment Please read this manual carefully before using the equipment be familiar with the operation and safety instructions of the equi...

Page 5: ...nteed Emergency handling If an emergency is found it shall be handled as follows Turn off the power switch Unplug Product responsibilities Without the written permission of the manufacturer it shall n...

Page 6: ...r oils and powders such as petroleum jelly and agar This sterilizer cannot be used for sterilization of liquids or liquids sealed with closed containers especially glassware which may cause the contai...

Page 7: ...he abnormal conditions chapter of the manual Necessary monitoring shall be carried out in accordance with relevant national and regional regulations After putting the monitoring substance such as chem...

Page 8: ...tomatic protection ensure the effectiveness of sterilization and disinfection 1 1 Product categories The product is classified as class I type B equipment according to electrical safety The product is...

Page 9: ...water vapor is between 134 and 137 and the temperature difference between each point in the sterilization chamber at the same time is no more than 2 and the maintenance time is 4min 2 sterilization f...

Page 10: ...rameter of autoclave Max working temperature 137 Max working pressure 0 24MPa Safety valve set pressure 0 24MPa Safety valve open pressure 0 24MPa 0 26MPa Capacity of clean water tank 1L Waste water t...

Page 11: ...heavy metals 10mg L 1mg L 0 2mg L 0 005mg L 0 05mg L 0 1mg L Chloride Phosphate Conductivity PH value Conductivity hardness 2mg L 0 5mg L 15uS cm 5 7 5 Colorless clean without precipitation 0 02mmol L...

Page 12: ...hether it meets the requirements according to the following clauses 3 2 1 Power requirements AC220V 50Hz unidirectional power supply voltage fluctuation does not exceed 10 power 3 5kVA 3 2 2 Environme...

Page 13: ...ich the sterilizer is placed is strong enough do not block the vent of the sterilizer 3 3 Handling of packaged products To ensure safe handling please use a cart and similar tools to move the device I...

Page 14: ...t of the product it is placed on the workbench Please note the following when handling It is strictly forbidden to lift the door during transportation to avoid damage to the device Do not lift the foo...

Page 15: ...rain Connect the pipe from the outlet of waste water chapter 1 4 to the sewers or storage containers 3 8 Insert the power plug into the socket Make sure that the equipment is reliably grounded 3 9 Oth...

Page 16: ...k test The leak test is done to ensure the tightness of the sterilization chamber and its pipelines to ensure that the cold air in the sterilization chamber is eliminated This device comes with a leak...

Page 17: ...aper plastic packaging or double layer packaging may not dry should be used immediately after sterilization Universal Non wrapped load or wrapped solid fabric and cavity load 134 solid Solid load with...

Page 18: ...No GC JS 21 Version 01 00 Effective date 2020 07 22 18 min tion mL min tion mL min tion mL min tion mL LFSS0 3AA 2 2kg 20 280 24 300 38 380 18 240...

Page 19: ...water quality 29 means that the current water quality is 29uS cm Shows whether the pre heating function is on On is on Off is not on Count 00001 indicates that it has been used once Display the curre...

Page 20: ...ation and other information please refer to Chapter 5 of this manual 6 1 3 Settings You can select the setting menu on the main page and press OK to enter The program menu after entering is as The pro...

Page 21: ...can press Start Stop to move to Month and so on after setting all dates and times press OK to save and exit 6 1 3 3 to switch different temperature units When cursor move to this option press OK butt...

Page 22: ...djust to higher drying class to solve this issue When cursor move to this option press OK button to adjust the drying class I II III The drying time corresponding to each level is as follows 134 Unive...

Page 23: ...he program once and water needs to be refilled 2 Need to close the door before running the program 3 Door is closed 4 The door is locked to ensure that the door cannot be opened during operation 5 Bad...

Page 24: ...representative is required to maintain the equipment 10 Last cycel ended improperly Since the last sterilization cycle was stopped by the user the power failure or failure did not end normally the loa...

Page 25: ...ing The equipment cannot run the program when the clean water tank is short of water and the water needs to be added to the water clean tank When there is no water in the equipment storage tank or whe...

Page 26: ...ponding to each program see chapter 5 2 1 There is no difference in sterilization effect between the 134 C sterilization procedure and the 121 C sterilization procedure It is recommended to to use the...

Page 27: ...you can turn the door handle counterclockwise to open the door 7 9 2 Offload After opening the door after sterilization the load can be taken out The operator should use the configured instrument tray...

Page 28: ...erilization chamber is completely released 1 Use a tool to open the safety valve to completely release the internal pressure Be careful not to get burned by the discharged water vapor Can be operated...

Page 29: ...32 Heating ring temperature 220 heating jacket temperature sensor damaged Check the temperature sensor in heating jacket 3 E33 Steam generator heating tube temperature 230 The temperature sensor of st...

Page 30: ...ed Clean and sterilize indoor filters Check the drain solenoid valve vacuum selection valve 14 E6 The door switch is disconnected when the program is running The door switch sensor is displaced the mi...

Page 31: ...eal 1 check if same result with empty loading 2 Check the vacuum pump 3 Check whether the door is closed enough to adjust the position of the door 21 E16 There is still a large negative pressure in th...

Page 32: ...cessories 9 2 Maintenance plan 9 2 1 Cleaning plan No Cleaning items Cycle Requirement Remark 1 Sterilization chamber cleaning once a week Keep clean without sewage Refer to 9 3 1 2 Clean water tank O...

Page 33: ...ine cleaning agents to clean the sterilization chamber Before cleaning the sterilization chamber please confirm that the sterilization room has been cooled to avoid burns 9 3 2 Cleaning the water stor...

Page 34: ...sical program 2 Wait until the pressure is near 100kpa pull the pull ring on the safety valve to make it open wait for about 1s there will be air and steam exhausted indicating that the safety valve c...

Page 35: ...it fails please contact the supplier or supplier representative The test results may be inaccurate when the sterilization room is hot or not dry and must be tested when the sterilization room is cold...

Page 36: ...eft and right as shown in the figure below After softening press the remaining part into the sealing groove at a time and then insert the cover plate and tighten the four screws When replacing the sea...

Page 37: ...itions are not covered by the free warranty even during the warranty period 1 Failure or damage caused by improper installation and use 2 Failure or damage caused by accidental drop or collision 3 Fai...

Page 38: ...terilized items Place on the instrument rack 2 Instrument rack optional Sterilized items in the lower layer Upper instrument tray 3 External waste water tank optional To store the waste water generate...

Page 39: ...Pass Empty load Inspection of finished product Pass Solid load Type test Pass Small amount of porous permeable strips Type test Pass Small amount of porous permeability load Type test Pass Fully poro...

Page 40: ...02 RAISE 15 53 00 021 089 2 00 15 VACUUM 15 53 50 080 095 2 00 50 RAISE 15 54 05 010 103 9 00 14 VACUUM 15 55 09 080 105 7 01 05 RAISE 15 56 24 221 133 7 01 15 STER START 15 56 24 221 133 7 01 15 MAX...

Page 41: ...injected into the sterilization chamber temperature and pressure will increase 5 drain The sterilization room is draining 6 Sterilization Sterilizing maintain sterilization temperature and pressure 7...

Reviews: