6.1.4.2 The program runs after pressing the "Start" button, the curve will flish, The total running time will start
counting.
The line on the process curve diagram will flash to indicate the running stage.
For the description of each stage, see Appendix III: Description of each stage
6.1.8 Explanation of each prompt message
No.
Prompt massages
Description
1
The water in the water storage tank is not enough to run the
program once, and water needs to be refilled
2
Need to close the door before running the program
3
Door is closed
4
The door is locked to ensure that the door cannot be opened
during operation
5
Bad water quality
The water in the water storage tank is unqualified: it needs
to be replaced with distilled or deionized water
6
Please open the door, wait machine to
cool down then try again
The sterilization chamber was too hot when the 134 ° C
program was just run, and then the 121 ° C program was run.
The door needs to be opened for cooling. This time takes up
to 5-7 minutes
7
Please replace bacterial filter
The life of the bacteria filter has expired and needs to be