No.:GC-JS-21 Version: 01/00
Effective date
:
2020-07-22
22
When cursor
“
”
move to this option
,
press
“
OK
”
button to adjust the setting
:
When set to "On": The equipment will start to warm up and keep warm after being powered on, in order to
shorten the sterilization cycle time;
When set to "off": the device will not preheat after power on, the device will only start to preheat after the
program is running, the entire sterilization cycle time will be 5-7 minutes longer than the "on"
It is recommended to set this setting to "On".
When this setting is turned on, special care should be taken not to touch the sterilization
chamber when the door is open to avoid burns.
6.1.3.6
:
The user can change or set the drying time of 4 programs. If the preset daying is
not good after sterilization, the user can adjust to higher drying class to solve this issue.
When cursor
“
”
move to this option
,
press
“
OK
”
button to adjust the drying class
:
I/II/III.
The drying time corresponding to each level is as follows
:
134
℃ Universal
134
℃ Universal
quick
134
℃ solid
121
℃ Universal
Class I
6
4
2
8
Class II
10
6
2
12
Class III
14
8
2
16
6.1.3.7
:
The user can set whether to enter with a password when starting
up, and set the entry password
:
When cursor
“
”
move to this option
,
press
“
OK
”
button to enter, interface as below:
6.1.4 To-be-run page and running interface
6.1.4.1 After the program is selected, enter the interface to be run, the interface is as follows
Move the cursor
“
”
to the "Password" field, and
press the "OK" key to enter the setting password:
Press the "Up" key once: 1
Press the "Down" key once: number -1
Press the "Start/Stop" button once: digital shift
After setting the required password, press "OK" to save
Move the cursor
“
”
to the "Enable" field and press
"OK" to set whether to enable:
ON: enabled; Off: not enabled