. Instructions of operation
7.1 Preparation before operation
Check the following items before operation
:
1
)
If the SD card is needed to save records, check whether the SD card is inserted
;
2
)
Check whether there is any impurity adsorption on the sealing surface of the door seal and the sealing
surface of the chamber, if it is clean
。
Can be powered on after checking.
7.2 Connect the waste water draining tube
Two ways
:
1)
Directly connected to the sewer; 2) Connected to the waste water storage container;
Connect the draining tube with connector to the waste water outlet of the device , and connect the other end to
the sewer or waste water storage container.
If it is connected to the waste water tank storage container, please check whether the
water storage container is full once a day. If it is full or is about to be full, please drain the waste
water storage container.
7.2 Waste water tank draining
When the waste water tank is full, you need to drain the waste water tank. When the waste water tank is full,
the program cannot run. Please drain the entire waste water tank when draining.
7.3 Clean Water tank filling
The equipment cannot run the program when the clean water tank is short of water, and the water needs to be
added to the water clean tank
:
When there is no water in the equipment storage tank or when there is a shortage of water, you need to add water
to the storage tank, at least 300ml.
7.4 Loading the instruments
Use the tray handle to load the instruments to avoid the hazard of scalding caused by touching the chamber.
Do not touch the wall of the sterilization chamber to avoid burns.
Pay attention when loading the instruments
:
1) The load is forbidden to contact the chamber wall, and there should be a gap between the sterilization
packages to facilitate the penetration of steam;
2) Distortion of the pipeline load is prohibited, and the pipeline should be kept smooth;
3) If it is an open type load, the opening should be facing down to prevent non-drying caused by water storage;
4) The instruments should be placed evenly, arranged at intervals and not allowed to be stacked, otherwise it
may cause insufficient sterilization and drying;
5) It is advisable to sterilize instruments, utensils or articles of the same material for the same batch. When the
materials are not the same, the textile is the upper layer, and the metal instruments are the lower layer;
6) When sterilizing instruments with paper-plastic packaging, it is recommended to put the paper-plastic