No.:GC-JS-21 Version: 01/00
Effective date
:
2020-07-22
9
1.4 Sterilization principle, main sterilization factors and strength
1.4.1 Sterilization principle
This equipment is suitable for the vacuum pump to exclude the cold air in the sterilizing room,
take saturated steam as the sterilizing factor, and use the characteristics of high latent heat and high
penetrability of saturated steam to sterilize the equipment.
1.4.2 Sterilization factor and strength
1) sterilization factor and strength of the sterilization procedure at 134
℃
: the temperature of
saturated water vapor is between 134
℃
and 137
℃
, and the temperature difference between each
point in the sterilization chamber at the same time is no more than 2
℃
, and the maintenance time is
4min;
2) sterilization factor and strength of the sterilization process at 121
℃
: the temperature of
saturated water vapor is between 121
℃
and 124
℃
, and the temperature difference between each
point in the sterilization chamber at the same time is no more than 2
℃
, and the maintenance time is
20min.