No.:GC-JS-21 Version: 01/00
Effective date
:
2020-07-22
27
Press "Start" to run the program automatically.
7.8 Man-made interruption
If the device needs to be stopped running during the cycle, press and hold the "Stop" key for 5 seconds, the
device will automatically release pressure to a safe pressure and end the program, then door can be opened.
7.9 Open the door and unload the loadings
7.9.1 After the program ends, you can turn the door handle counterclockwise to open the door
;
7.9.2 Offload
After opening the door after sterilization, the load can be taken out. The operator should use the configured
instrument tray handle or wear insulated gloves or other tools to remove the sterilized items.
The operator should use the configured instrument tray handle or wear insulated gloves or
other tools to remove the sterilized items to avoid burns.
If the sterilized items are taken out immediately after the procedure, there may be a
small amount of condensate on the sterilized items, and the door should be opened and wait for 5