78
4.3.3 Min/Max
Definition
When the Min/Max function is enabled, the multimeter takes in a series
of readings from the measurements, stores the minimum and maximum
readings in the memory, and then calculates the average value of all
readings. The number of readings taken since Min/Max operation is
enabled is recorded as well. These data are stored in a volatile memory
and will be cleared when the meter is turned off, or when the MIN/MAX is
turned on, or after a remote interface reset. The meter beeps every time
when it captures a new minimum or maximum.
How to use Min/Max
You can use the Min/Max feature either through the front panel operation
or the remote interface operation.
Front Panel Operation
User first selects the measurement function they wish to use, and then
press MIN/MAX button to enable the Min/Max function. To read the
min/max value, the average value or the readings count, user can press
CONFIG button and then MIN/MAX button. Use
◁
and
▷
buttons to
switch between those values. The locations of the buttons are shown
with red rectangle frames in
Figure 4-16
.
※
Note: Press MIN/MAX again to disable this feature. The “MATH”
anunnciator on the display indicates the state of a mathematical feature.
Figure 4-16
Summary of Contents for M3500A
Page 1: ...An ISO 9001 2000 Company 6 Digit Multimeter M3500A Operation Manual ...
Page 2: ...Operation Manual M3500A DMM ...
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 10: ......
Page 11: ......
Page 26: ...1 2 3 4 5 6 ...
Page 33: ......
Page 34: ......
Page 35: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 42: ......
Page 43: ......
Page 44: ......
Page 45: ......
Page 47: ......
Page 48: ...48 Figure 3 18 5 ...
Page 95: ......
Page 104: ......
Page 151: ......
Page 155: ......