145
variable with at least 35 characters).
Other Interface Commands
SYSTem:LOCal
Place the multimeter in the
local
mode. All buttons on the front panel are fully
functional.
SYSTem:REMote
Place the multimeter in the
remote
mode. All buttons on the front panel, except
the LOCAL button, are disabled.
B.9 Status Reporting Commands
SYSTem:ERRor?
Query the multimeter’s error queue. Up to 20 errors can be stored in the
queue. Errors are retrieved in first-in-first out (FIFO) order. Each error
string may contain up to 80 characters.
STATus:QUEStionable:ENABle <enable value>
Enable bits in the Questionable Data enable register. The selected bits are then
reported to the Status Byte.
STATus:QUEStionable:ENABle?
Query the Questionable Data enable register. The multimeter returns a
binary-weighted decimal representing the bits set in the enable register.
STATus:QUEStionable:EVENt?
Query the Questionable Data event register. The multimeter returns a
decimal value which corresponds to the binary-weighted sum of all bits
set in the register.
STATus:PRESet
Clear all bits in the Questionable Data enable register.
*CLS
Clear the Status Byte summary register and all event registers.
Summary of Contents for M3500A
Page 1: ...An ISO 9001 2000 Company 6 Digit Multimeter M3500A Operation Manual ...
Page 2: ...Operation Manual M3500A DMM ...
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 10: ......
Page 11: ......
Page 26: ...1 2 3 4 5 6 ...
Page 33: ......
Page 34: ......
Page 35: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 42: ......
Page 43: ......
Page 44: ......
Page 45: ......
Page 47: ......
Page 48: ...48 Figure 3 18 5 ...
Page 95: ......
Page 104: ......
Page 151: ......
Page 155: ......