4.2.1 Trigger Mode
There are three trigger modes in M3500A: auto, immediate, and single
triggering. User can specify the trigger mode for their measurement. The
factory default is auto triggering when the meter is power-on.
A. Auto Triggering Mode (Front Panel Operation only)
Definition
Auto triggering takes continuous readings at a fastest rate possible for
the present measurement. The rate of taking readings depends on the
current settings. This function is only available through the front panel.
The auto triggering is also the default for trigger mode in M3500A.
How to use Auto Trigger
Press ATUO TRIGGER on the front panel to toggle for enabling auto
trigger mode. The location of the button is shown with a red rectangle
frame on
Figure 4-9
.
Figure 4-9
B. Internal Immediate Trigger Mode (Remote Interface Operation Only)
Definition
The immediate triggering mode is only available through the remote
interface operations and it is the default for the remote interface
operations. When the meter is in immediate trigger mode, a trigger will
be issued immediately as soon as the meter is in “wait for event” state.
Summary of Contents for M3500A
Page 1: ...An ISO 9001 2000 Company 6 Digit Multimeter M3500A Operation Manual ...
Page 2: ...Operation Manual M3500A DMM ...
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 10: ......
Page 11: ......
Page 26: ...1 2 3 4 5 6 ...
Page 33: ......
Page 34: ......
Page 35: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 42: ......
Page 43: ......
Page 44: ......
Page 45: ......
Page 47: ......
Page 48: ...48 Figure 3 18 5 ...
Page 95: ......
Page 104: ......
Page 151: ......
Page 155: ......