51
Fast 5 ½ digits
Off
Off
0.1
Slow 5 ½ digits
On
On
1
Fast 6 ½ digits
On
On
1
Slow 6 ½ digits
On
On
10
Table 4-1
※
Note: To change resolution, refer to Section 4.1.3: Resolution Setting
in this chapter.
Figure 4-1
Remote Interface Operation
To set Auto Zero and Auto Gain through the remote interface, use the
following commands:
SENSe:ZERO:AUTO {OFF|ONCE|ON}
SENSe:GAIN:AUTO {OFF|ONCE|ON}
The OFF and ONCE parameters have a similar effect. Auto Zero OFF or
Auto Gain OFF doesn’t issue a new offset measurement. However, Auto
Zero ONCE or Auto Gain ONCE issues an immediate offset measurement.