11.3 Adjusting 50mA DC Current
Step 1:
Make sure the lead cables are not connected to the standard source.
Step 2:
Using the
〔
mA
〕
key, select DC current function.
Step 3:
Connect the lead cables to the output terminals of the standard source as shown in Figure 17.
Step 4:
The measured value and unit shall be shown in the upper part of the LCD.
Step 5:
Pressing the
〔
ZERO
〕
key enters the 50mA of DCmA
CAL mode. The LCD shows the present adjusting point
“CAL:0” in the lower part and the reference current and unit
needed for the point in the upper part.
Step 6:
Pressing the
〔
〕
key saves the adjusted value
and the LCD shows “SAVE” symbol in the upper
part for 2 seconds.
Step 7:
Pressing the
〔
ZERO
〕
key exits the CAL mode. The
adjustment is finished. Figure 17 Calibrating DC current 50mA
standard source
(5520A)