30
Step 8:
To re-enter into the frequency set mode, press the
〔
FREQ
〕
key to set the frequency.
Step 9:
Pressing the
〔
ON
〕
key causes the SOURCE indicator on the LCD to change from ‘’OFF‘’ to ‘’ON ‘’, and
the calibrator sources low level between the output terminals.
Step 10:
Pressing the
〔
START
〕
key the calibrator sources the set number of pulse responding to the preset
frequency and amplitude, LCD shows the symbol “RUN”.
Step 11:
When source is complete, the calibrator automatically turns off the output and ceases operation. The
“RUN” symbol disappears from the LCD.
Step 12:
To turn off the output, press the
〔
ON
〕
key once again. The ‘’OFF‘’ appears on the LCD and no signals
sourced between the terminals.
Tips:
z
The frequency range of the pulse could only be changed by pressing
〔
RANGE
〕
key in the frequency set mode.
z
In the pulse sourcing process, pressing the
〔
START
〕
key causes to stop the output, and the “RUN” mark
vanishes from the LCD. Press the
〔
START
〕
key once more to restart the sourcing function.
z
Restarting the pulse output requires the source function is in ‘’ON ‘’state.