Tips:
z
If the battery level is below 25% full, the adjustment operation can’t be operated. And the LCD shows
“ERR” in the lower part.
11.1 Adjusting all ranges of the DC Voltage
Step 1:
Make sure the lead cables for measurement are
not connected to the measuring instrument under test.
Step 2:
Using the
〔
V
〕
key, select DC voltage function.
Step 3:
Connect the lead cables to the output terminals of
the standard source as shown in Figure16.
Step 4
:Pressing the
〔
RANGE
〕
key selects the range. The
measured value and unit shall be shown in the upper part Figure 16 Calibrating DC voltage and frequency
of the LCD.
Step 5:
Pressing the
〔
ZERO
〕
key enters the measurement CAL mode. The LCD shows the present adjusting
point “CAL:0” in the lower part and the reference voltage and unit needed for the point in the upper part.
TRIG
OUT
Lo
NORMAL
SCOPE
Hi
AUX
standard source