Manual OPTISENS AAM 1050
OPTISENS AAM 1050
24
9.
Output
Main menu
.
.
Basic settings
Basic settings
.
.
Analog output
Analog output
Select output
Selection;
.
0/4 – 20 mA
- between 0 – 20 mA
and 4 – 20 mA
.
Begin 0/4 mA
- Measured value at
4 mA
.
End 20 mA
- Measured value
at 20 mA
Analog output
- between readout of the
Measuring value
Measurement or the
Control values
Serial interface
RS485
Bus address
Bus address
Nr. 00
Set a number
between 0 and 31
to address the
device via data
bus.
9.1
Current output
You can read out the measured values and the temperature as 0/4…20 mA signals via the current outputs.
With the setting 4…20 mA the resolution is lower, but defective cable connections are immediately evident.
With the parameters Begin and End you define which part of the measuring range you want to read out.
Alternatively you can use the current output as steady-state controller output. In that case assign the current
output to controller S1 or S2, respectively.
9.2
Serial interface RS485
The devices are available with serial interface RS485 by means of which they can be integrated in a data
bus system. Via the interface, all settings, measured and control values as well as any error messages can
be read out digitally.
Devices with RS485 are automatically delivered with the leaflet „Information on the RS485“ which contains
instructions on the communication and a complete list of the functions available via interface.