Manual OPTISENS AAM 1050
OPTISENS AAM 1050
12
6.1
Configuration of the Converter
The instruments AAC 050 provide two measuring inputs.
Input 1 can work with the following sensors:
Parameter
Measuring range
Free Chlorine
Cl2
0.00…4.00 mg/l
Free Chlorine*
Cl2
0.00…20.00 mg/l*
Chlorine dioxide
ClO2
0.00…4.00 mg/l
Ozone
O3
0.00…4.00 mg/l
Hydrogen peroxide
H2O2
0.00…30.00 mg/l
* special version
Input 2 works with the temperature sensors.
NOTE
The configuration of measuring parameter, measuring range and type of temperature sensor is done at the
factory and cannot be changed after shipping!
6.2
Calibration
Calibration of the measurement of Cl2, ClO2, O3 and H2O2
Main menu
Calibration
DPD Calibration
DPD Calibration
0.48 mg/l
Slope
Enter the concentration determined by
photometric measurement as calibration
value.
025 mV
1) Switch off the controller. Take a sample of the test water flowing out of the armature and determine the
concentration by photometric DPD measurement or a similar reference method.
2) Enter the determined concentration and calibrate by pressing keys
and
:
Start with key
and then -
while still applying pressure on this key - additionally press key
.
3) Check the displayed slope before switching on the controller.