Manual OPTISENS AAM 1050
5
OPTISENS AAM 1050
0.4
Features
0.4.1
Device
0.00…4.00 mg/l
Cl2
free Chlorine
0.00…20.00 mg/l
Cl2
free Chlorine *
0.00…4.00 mg/l
O3
Ozone*
0.00…4.00 mg/l
ClO2
Chlorine dioxide*
Measuring ranges
0.0…30.00 mg/l
H2O2
Hydrogen peroxide*
*special version
Temperature range
-30.0…+140.0°C
Minimum flow rate
> 30.0 l/h
Display
measured value with unit;
temperature with unit;
status display sensor, calibration, controller & alarm
Temperature compensation
manual or automatic
Calibration
1-point calibration;
Zero-point calibration for low concentrations
Automatic Sensor Cleaning
Automatic recognition
Flow monitor
Switch point at 30 l/h
The configuration of measuring parameter, measuring range and type of temperature sensor is done at the
factory and cannot be changed after shipping!
0.4.2
Controller
Set points
2 set points free adjustable within measuring range with choice of
direction
Controller types
ON/OFF controller with hysteresis
P controller as Pulse-Pause-, Impulse-Frequency- or steady controller
PI controller as Pulse-Pause-, Impulse-Frequency- or steady controller
Hysteresis
0.00 – 4.00 mg/l or 0.0-30.0 mg/l for H2O2
P range X
P
0.00 – 4.00 mg/l or 0.0-30.0 mg/l for H2O2
Integral time T
N
0…2000 s
Least pulse
0.1…9.9 s
Pulse+ pause time
02…99 s
Impulse frequency
00…72 equiv. to 0...7200 pulses/h
Turn-on delay
0…200 s
Alarm function
min. and max. limit and onset delay, power failure alarm,
Dosage control
0… 90 min
0.4.3
Connections
Relays
3 potential-free contacts (2x controller, 1x alarm)
6 A, 250 V, max. 550 VA
Analog output
1x 0/4…20 mA galvanically isolated
max. loading 500 Ohm
Analog inputs
1 measuring input for Cl2, ClO2, O3 or H2O2,
1 measuring input for Pt100
Digital input
external controller stop and/or lack-of-water indication
Serial interface
RS485, Baud rate 9600, data format 8 Bit,
1start and 1stop bit, no parity
Water Inlet/Outlet
Female thread ½“ with attached valve,
hose connection 6 mm I.D.,8 mm O.D.
Flow required 10-300 l/h, ideal 50-200 l/h, pressure max. 6 bar
Sampling point