3
INSTALLATION
22
OPTIMASS 2000
www.krohne.com
08/2014 - 4000810003- MA OPTIMASS 2000 R03 en
Heating times
Reference conditions
3.4.10 Purge ports
If the meter has been ordered with a purge port, it will be supplied with NPT female connections
which are clearly marked. The connections are sealed with NPT plugs and PTFE tape.
3.4.11 Burst discs
If the meter has been ordered with a burst disc, it will be supplied with the disc fitted. The failure
pressure of the disc is 20 barg at +20
°
C / 290 psig at +68
°
F.
Automatic fitment
Automatic fitment
Automatic fitment
Automatic fitment
If the connection rating of the meter is greater than 100 barg / 1450 psig but the 150 barg / 2175
psig secondary containment option has not been purchased, the meter will be supplied with a
burst disc fitted as a safety feature.
CAUTION!
The maximum heating pressure and temperature for heating jackets is 10 barg at
130
°
C / 145 psig at 266
°
F
Temperature [
°
C /
°
F]
1
Time [minutes]
S100
S150
S250
40 / 104
7
7
10
60 / 140
10
10
17
80 / 176
15
15
30
100 / 212
20
20
60
110 / 230
30
50
90
120 / 248
75
200
270
1
Measured at the spigot end of the measuring tube.
Ambient temperature
+25
°
C / +77
°
F
Heating medium
Hot liquid
Heating medium temperature
+130
°
C / +266
°
F
CAUTION!
DO NOT remove these plugs.
The meter is factory sealed with a dry nitrogen gas fill and if moisture is allowed to enter the
meter casing it will cause damage. The plugs should only be removed to purge the meter casing
in the event that the primary measuring tube fails.
If it is suspected that the primary measuring tube has failed, de-pressurise the meter and
remove it from service, as soon as it is safe to do so.