Installation and operating instructions
IFC 020
5/5
5.7 Pulse output P
Fct. 1.06 PULS.OUTP. P
Press key
→
.
→
FUNCT. P = select function for pulse output,
press
→
key
• OFF
(switched off, no function)
• 1 DIR.
(1 flow direction)
• 2 DIR.
(2 flow directions, F/R mode, forward/reverse)
Select with key
↑
or
↓
.
Press
↵
key to transfer to subfunction “SELECT P”.
Exception:
When “OFF” selected, return to Fct. 1.06 PULS P.
→
SELECT P = select pulse type,
press
→
key
• 100 Hz
• 1000 Hz
• PULSE/VOL.
(pulses per unit volume, flow)
• PULSE/ZEIT
(pulses per unit time for 100% flow)
Select using
↑
or
↓
key.
Transfer to subfunction ”PULSWIDTH” with
↵
key.
Note:
when 100 Hz or 1000 Hz selected, return to Fct. 1.06 PULS.OUTP. P.
→
PULSWIDTH = set pulse width,
press
→
→
key
•
50 mSec
F
max
= 10 Hz
F
min
= 0.0056 Hz (= 20 Pulse / hr)
• 100 mSec
= 5 Hz
• 200 mSec
= 2.5 Hz
• 500 mSec
= 1 Hz
•
50
1
m
Sec
= 0,5 Hz
Select using
↑
or
↓
key.
Transfer to subfunction ”VALUE P” with
↵
key or return to Fct. 1.06
PULS.OUT. P, depending on choice of pulse type in subfunction ”SELECT P”.