Multicheck6010
EN
- 62 -
3.8. Help Screen
The Help screens contain diagrams that show the correct use of the device.
Figure 3 3: example of a help screen
Press the HLP key to enter the help screen
Press the HLP key or the Exit/Back/Return key to exit the help screen
Press the Left and Right keys to switch to previous/next help screen
4. Measurements
4.1 Insulation resistance
How to perform an insulation resistance measurement
Step 1
Select
Insulation
function with the function selector FCT key. The following menu is
displayed:
Figure 4 1: Insulation resistance measurement menu
Step 2
Set the following measuring parameter and limit values:
•
Volt:
Nominal test voltage
•
Límite:
Low limit resistance value