33
Scanning the Object
High Quality Mode
The instrument emits a laser beam on the object, and captures the rays reflected from the object using the CCD
for measurement.
Thus, for better measurement accuracy, it is necessary for the intensity of the reflected rays to be maintained
within the CCD’s measurement range (i.e. the range that ensures linearity).
In high quality mode, the parts of the reflected rays that are outside the CCD’s measurement range will be elim-
inated, so that only the data with assured linearity is output.
This provides measured data with higher accuracy than that which includes reflected rays outside the linearity
range.
M
emo
If the instrument is set to High Quality Mode when you read in memory-card data that was scanned at a VIVID 910, the
instrument will automatically eliminate all data values that are outside the measurement range. This will be true even if
High Quality Mode was not turned on at the time the original data were scanned.
N
ote
When reading in memory-card data that was scanned at a VIVID 900, the instrument will not eliminate data values
that are outside the measurement range.
[Operating Procedure]
1
Press .
The MENU view will appear.
2
Press
to locate the arrow cursor to
“CAMERA SETTINGS”.
3
Press .
The CAMERA SETTINGS view will appear.
4
Press
to locate the arrow cursor to
“HIGH QUALITY MODE”.
5
Press .
The currently selected high quality mode (ON or OFF)
will turn to pink.
6
Press
to select “ON” or “OFF”.
<Settings>
•
ON:
High quality mode
•
OFF:
Digitize mode (normal measurement mode)
Vivid910-hard-E.book Page 33 Thursday, September 25, 2003 6:49 PM
Summary of Contents for Vivid 910
Page 13: ...11 Chapter 2 Preparation Vivid910 hard E book Page 11 Thursday September 25 2003 6 49 PM...
Page 27: ...25 Chapter 3 Operation Vivid910 hard E book Page 25 Thursday September 25 2003 6 49 PM...
Page 55: ...53 Chapter 4 Appendix Vivid910 hard E book Page 53 Thursday September 25 2003 6 49 PM...
Page 63: ...Vivid910 hard E hyou fm Page 3 Thursday September 25 2003 6 52 PM...
Page 64: ...2001 2006 KONICA MINOLTA SENSING INC AGKCKK 7 Printed in Japan 9222 1888 40 8KXKF JCTF J QW HO...