Maintenance Mode
3
The following message appears in the display window�
4
Use the [Test print] and [Calibration] buttons to move the cursor to an
item, and press the [Service/Enter] button to select and set that item.
• [QC History]
Press the [Service/Enter] button to display the QC history screen.
See "QC History Screen".
• [Clear QC History]
Press the [Service/Enter] button to display the Clear QC history screen.
See "Clear QC History Screen".
5
Press the [Reset/Cancel] button to return to the previous screen.
Results Data Screen
This displays the results for the three most recent QC patterns.
The density values (x100) for steps 1 to 4 are displayed.
If there is a density read error for steps 1 to 4, [NG] appears.
(If [NG] appears, perform a calibration, and then try printing the QC pattern again.)
Results data
Selected tray
The displayed results data is rounded off to after the decimal point.
Therefore, the results may be different when the same values are
displayed�
•
Use the [Test print] and [Calibration] buttons to change the selected tray.
• Alternatively, press the tray button to change the selected tray.
• Press the [Reset/Cancel] button to return to the QC results screen.
QC Results
Clear QC History
Summary of Contents for DRYPRO 873
Page 1: ...EN 13...
Page 2: ......
Page 9: ...9 Chapter 1 Safety Warnings Precautions...
Page 25: ...25 Chapter 2 Product Outline...
Page 35: ...35 Chapter 3 Operation from Main Unit...
Page 76: ...76...
Page 77: ...77 Chapter 4 Troubleshooting...
Page 100: ...100...
Page 101: ...101 Chapter 5 Maintenance...
Page 115: ...115 Appendix...
Page 120: ...120...
Page 121: ...121...
Page 122: ...122...
Page 123: ......