Safety Precautions
Table 3 (support for IEC 60601-1-2:2007)
Guidance and manufacturer’s declaration - electromagnetic immunity
The DRYPRO 873 is intended for use in the electromagnetic environment specified below. The customer or the
user or the DRYPRO 873 should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vemf 150 kHz
to 80 MHz
3V/m 80 MHz to
2.5 GHz
[3]V
[3]V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the DRYPRO
873, including cables, than the recommended separa-
tion distance calculated from the equation applicable
to the frequency of the transmitter.Recommended
separation distance
d=[1.2] √ P
d=[1.2] √ P80 MHz to 800 MHz
d=[2.3] √ P800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m)
Field strengths from fixed RF transmitters, as de-
termined by an electromagnetic site survey,
a
should
be less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
[NOTE] At 80 MHz and 800 MHz, the higher frequency range applies.
[NOTE] These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theo-
retically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electro-
magnetic site survey should be considered. If the measured field strength in the location in which the DRY-
PRO 873 is used exceeds the applicable RF compliance level above, the DRYPRO 873 should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re- orienting or relocation the DRYPRO 873.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [3]V/m.
Summary of Contents for DRYPRO 873
Page 1: ...EN 13...
Page 2: ......
Page 9: ...9 Chapter 1 Safety Warnings Precautions...
Page 25: ...25 Chapter 2 Product Outline...
Page 35: ...35 Chapter 3 Operation from Main Unit...
Page 76: ...76...
Page 77: ...77 Chapter 4 Troubleshooting...
Page 100: ...100...
Page 101: ...101 Chapter 5 Maintenance...
Page 115: ...115 Appendix...
Page 120: ...120...
Page 121: ...121...
Page 122: ...122...
Page 123: ......