MIM-
MIM 01/0419
page
45
14. Technical Information
Measurement process:
electromagnetic
Range:
see order details
Media:
conductive
fluids
Minimum conductivity:
≥
20 µS/cm
Max. medium viscosity:
70 mm2/s
Max. pressure:
16 bar
Accuracy:
<±(0.8% of 0.5% of full scale)*
Repeatability:
±0.2% of full scale
Temperature
Measurement of media:
PT1000, range -20 °C …+70 °C
Response time flow t90
(alarm output /pulse output):
<250 ms
Response time temperature t90
(signal output):
<20 s
Mounting position:
in all directions
In-/outlet:
3 x DN/2 x DN
Pressure drop (max at ME)
MB 05
400 mbar
MB 10
50 mbar
MB 15
250 mbar (3/4 "), 30mbar (1")
MB 20
80 mbar
Handling:
4 optical touch fields,
useable
with
hand
gloves
Housing:
stainless
steel
1.4404,
display
screen
PMMA
Wetted parts
Connection fitting and
housing:
stainless steel 1.4404 (316 SS)
Insulation parts:
PEEK
Electrodes:
stainless steel 1.4404 (316 SS)
Seals:
FKM
Protection:
IP
67
Media temperature:
-20 °C … +70 °C
Ambient temperature:
-20 °C … +60 °C
Electrical data
Supply voltage:
19 - 30 VDC, internal power consumption
max. 200 mA (without outputs)
Display:
TFT display, 128 x 128 pixels,
1.4" display orientation in 90° steps
adjustable
Display repetition rate:
0.5 ... 10 s, adjustable
Pulse output
Push-Pull, freely scalable, configurable for
partial or accumulated totalizer