11. Status
The electromagnetic flowmeter can detect and display various device or application
errors.
If there is a status or error message, the STATUS symbol in the display alternately
flashes orange / red. To call up the status / error information, the status key must
be pressed, then the status window that appears then lists all the messages that
have accumulated up to this point in time. By pressing the
key, the user
confirms the knowledge of the displayed errors, the status memory is cleared and
the status window is closed. If one of the displayed errors persists, it will be reported
again by flashing the status icon.
The following status / error messages are generated:
Display text
Description
Debugging
Empty Pipe
Measuring tube is not completely
filled with medium or medium
with too low conductivity is used.
Check the filling of the
measuring circuit or
conductivity of the medium
(> 20
μ
s / cm)
Temp Sens Error
Error in the temperature
measuring circuit
Repair by KOBOLD Service
necessary
Meas saturated
Flow measuring circuit
overdriven
Reduce flow rate
No Subslave
Internal hardware error
Repair by KOBOLD Service
necessary
Simulation
Simulation function active
-