MIM-
MIM 01/0419
page
51
17.4 IO Link ISDU parameter table
ISDU
[hex]
Name
Description
Factory
default
value
Max
value
Min
value
Length
(Bytes)
Data
type
Access
Product
Identification
(Vendor
specific
parameters)
0x0010
VendorName
Kobold
Messring
max.
20
StringT
R
0x0011
VendorText
www.kobold.com
max.
32
StringT
R
0x0012
ProductName
MIM
‐
12xxxxxC3T0
max.
16
StringT
R
0x0013
ProductID
MIM
‐
12xxxxxC3T0
max.
16
StringT
R
0x0014
ProductText
magnetic
inductive
flow
meter
max.
32
StringT
R
0x0015
Serialnumber
xxxxxxx
max.
8
StringT
R
0x0016
HardwareRevision
Vxx.xx
max.
8
StringT
R
0x0017
FirmwareRevision
Vxx.xx
max.
8
StringT
R
0x0018
ApplicationDeviceTag
configurable
by
user
***
32
StringT
R/W
0x0019
FunctionTag
configurable
by
user
32
StringT
R/W
0x0020
LocationTag
configurable
by
user
32
StringT
R/W
Device
Status
Information
0x0024
DeviceStatus
0
‐
Device
OK
1
‐
Maintenance
required
2
‐
Out
of
specification
3
‐
Functional
check
4
‐
Failure
1
UIntegerT
R
0x0025
DetailedDeviceStatus
max.
20
ArrayT
of
OctetString
R
Display
Configuration
0x0100
DisplayOrientation
Orientation
of
display
1
(0)
‐
Landscape
(1)
‐
Portrait
Flip
(2)
‐
Landscape
Flip
(3)
‐
Portrait
1
UIntegerT
R/W
0x0103
DisplayLayout
Single
or
dual
layout
1
(0)
‐
single
(1)
‐
dual
1
UIntegerT
R/W
0x0104
UpperDisplay
Source
for
the
upper
display
0
(0)
‐
Flow
(1)
‐
Volume
(2)
‐
Temperature
(3)
‐
Part
Volume
1
UIntegerT
R/W
0x0105
LowerDisplay
Source
for
the
lower
display
2
(0)
‐
Flow
(1)
‐
Volume
(2)
‐
Temperature
(3)
‐
Part
Volume
1
UIntegerT
R/W
0x0106
DisplayRefresh
Time
Refresh
interval
for
the
display
[s]
0,5
10,0
0,5
4
FloatT
R/W
0x010A
LeftHotkey
Function
Function
for
left
hotkey
0
(0)
‐
Off
(1)
‐
Value
(2)
‐
MinMax
1
UIntegerT
R/W
0x010B
RightHotkey
Function
Function
for
right
hotkey
0
(0)
‐
Off
(1)
‐
Value
(2)
‐
MinMax
1
UIntegerT
R/W
0x010C
LeftHotkeySource
Source
for
the
left
hotkey
0
(0)
‐
Flow
(1)
‐
Volume
(2)
‐
Temperature
(3)
‐
Part
Volume
1
UIntegerT
R/W