![Knick Protos CONDI 3400(X)-051 User Manual Download Page 80](http://html.mh-extra.com/html/knick/protos-condi-3400-x-051/protos-condi-3400-x-051_user-manual_1980169080.webp)
80
Messages
CONDI 3400(X)-051 Module
No.
CONDI messages
Message type
T008
Meas. processing (factory settings)
FAIL
T009
Module failure (Firmware Flash check sum)
FAIL
T010
Conductivity range
FAIL / WARN
T011
Conductivity Alarm LO_LO
FAIL
T012
Conductivity Alarm LO
WARN
T013
Conductivity Alarm HI
WARN
T014
Conductivity Alarm HI_HI
FAIL
T015
Temperature range
FAIL
T016
Temperature Alarm LO_LO
FAIL
T017
Temperature Alarm LO
WARN
T018
Temperature Alarm HI
WARN
T019
Temperature Alarm HI_HI
FAIL
T020
Resistivity range
FAIL / WARN
T021
Resistivity Alarm LO_LO
FAIL
T022
Resistivity Alarm LO
WARN
T023
Resistivity Alarm HI
WARN
T024
Resistivity Alarm HI_HI
FAIL
T025
Concentration range
FAIL / WARN
T026
Concentration Alarm LO_LO
FAIL
T027
Concentration Alarm LO
WARN
T028
Concentration Alarm HI
WARN
T029
Concentration Alarm HI_HI
FAIL
T030
Zero range
WARN
T035
Cell factor range
WARN
T040
Salinity range
FAIL / WARN
T041
Salinity Alarm LO_LO
FAIL
T042
Salinity Alarm LO
WARN
T043
Salinity Alarm HI
WARN