Calibration / Adjustment
Data Entry of Premeasured Sensors
Data Entry of Premeasured Sensors
Entry of cell factor and zero point of a sensor, related to 25°C,
1013 mbars.
During calibration the module is in HOLD mode.
Current outputs and relay contacts of the module behave as configured
(Module BASE).
Menu Display
Data entry of
premeasured sensors
Select module: CONDI
During calibration, the output currents
(1 and 2), limit contacts, and controller
output are in HOLD mode.
Confirm with enter
Select calibration method
“Data entry”
Confirm with enter
Enter cell factor of premeasured
sensor.
Confirm with “OK”
or repeat calibration.
Calibration
Return
25.0°C
Module PH 3400-032
Info
1.225 mS/cm
Module OXY 3400-062
Module CONDI 3400-051
Module CONDI 3400-051
Return
25.0 °C
1.225 mS/cm
Product calibration
Zero correction
Automatic with standard cal solution
Manual entry of cal solution
Data entry
Temperature
25.6 °C
1.225 mS/cm
Conductivity
+25.6°C
1.225 mS/cm
Data entry - premeasured sensor
Cell factor
1
.980
End
Data entry of premeasured
sensors
Zero
+000.0 µS/cm