35
Menu Display
Product calibration
Select calibration menu
Select “Module CONDI”
Calibration method
“Product calibration”,
confirm with enter.
Step 1
Take sample.
Save measured value and tem-
perature at the moment of sampling
(“Save” softkey or enter).
The analyzer automatically returns to
calibration mode selection.
Press meas to return to measure-
ment.
Exception:
Sample value can be measured on
the site and be entered immediately.
To do so, press “Input” softkey.
Step 2
Lab value has been measured.
When you open the Product calibra-
tion menu again, the display shown
on the left appears: Enter reference
value (“Lab value”). Confirm with “OK”
or repeat calibration.
Module CONDI 3400-051
Return
25.0 °C
1.225 mS/cm
Data entry - premeasured sensor
Automatic with standard cal solution
Manual entry of cal solution
Product calibration
Product calibration
Cal medium: Product
Cal by taking sample
25.0 °C
1.225 mS/cm
Start
Return
and entering conductivity
Product calibration
Step 1: Sampling
“Save” the sample value
25.0 °C
1.225 mS/cm
Save
Input
“Input” lab value
Conductivity
Temperature
1.225 mS/cm
25.0 °C
Product calibration
Step 2: Lab value
Input sample lab value
25.0 °C
1.225 mS/cm
OK
Abort
Lab value
1
.222mS/cm
without temp compensation
Calibration data record
Calibration
25.6 °C
1.225 mS/cm
Cal mode
Cal temp
03/31/10 11:37
Cell factor
+025.6°C
Product calibration
Adjust
End
Adjustment
Press “Adjust” to take over the values
determined during calibration for
calculating the measured variables.
1.980
Sensor replacement
Zero correction